Getting Started
R&S
®
FPL1000
40
User Manual 1178.3370.02 ─ 11
4.2.2.8
Aux. Port
A 25-pole SUB-D male connector used as an input and output for low-voltage TTL con-
trol signals (max. 5
V). This connector is provided by the "Additional Interfaces" option
R&S
FPL1-B5.
13
25
1
14
Short-circuit hazard
Always observe the designated pin assignment. A short-circuit can damage the port.
Table 4-5: Pin assignment for optional AUX port
Pin
Signal
Description
1
GND
Ground
2
not used for spectrum analysis
3
GND
Ground
4
not used for spectrum analysis
5
GND
Ground
6
not used for spectrum analysis
7
GND
Ground
8 to 11
not used for spectrum analysis
12
GND
Ground
13
+5 V / max. 250 mA
Supply voltage for external circuits
14 to 19
I/O_<no.>
Control lines for user ports (see user manual)
20
GND
Ground
21
READY FOR TRIG-
GER
Signal indicating that the instrument is ready to receive a trigger
signal.
The signal polarity is configurable.
22 to 25
not used for spectrum analysis
23
GND
Ground
24 to 25
not used for spectrum analysis