The Spectrum Application (RF Measurements)
R&S
®
FPL1000
265
User Manual 1178.3370.02 ─ 11
"CISPR Aver-
age"
Determines a weighted average signal level according to CISPR
16-1-1.
The average value according to CISPR 16-1-1 is the maximum value
detected while calculating the linear average value during the speci-
fied dwell time.
The "CISPR Average" detector is only available for the CISPR filter.
"RMS Aver-
age"
A combination of the RMS detector (for pulse repetition frequencies
above a corner frequency) and the Average detector (for pulse repeti-
tion frequencies below the corner frequency).
Lowpass filters of the second order determine the average value
(simulation of a mechanical pointer instrument).
The "RMS Average" detector is only available for the CISPR filter.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DETector
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DETector
Select Marker
The "Select Marker" function opens a dialog box to select and activate or deactivate
one or more markers quickly.
Remote command:
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
EMI Final Measurement Configuration
The final EMI measurement can be performed with different settings than the initial
peak search. These settings are described here.
Measurements and Results