Remote Commands
R&S
®
FPL1000
974
User Manual 1178.3370.02 ─ 09
Example:
SYST:ERR:EXT? ALL
Returns all messages for the currently active application, e.g.
"Message 1","Message 2"
.
Example:
SYST:ERR:EXT? FAT,'Spectrum2'
Queries fatal errors in the 'Spectrum2' application. If none have
occurred, the result is:
"
"
.
Usage:
Query only
SYSTem:ERRor:LIST?
[<MessType>]
This command queries the error messages that occur during R&S
FPL1000 operation.
Query parameters:
<MessType>
SMSG | REMote
SMSG
(default) Queries the system messages which occurred during
manual operation.
REMote
Queries the error messages that occurred during remote opera-
tion.
Note: The remote error list is automatically cleared when the
R&S
FPL1000 is shut down.
Return values:
<SystemMessages> String containing all messages in the "System Messages" table.
<RemoteErrors>
<Error_no> | <Description> | <Command> | <Date> | <Time>
Comma-separated list of errors from the "Remote Errors" table,
where:
<Error_no>
: device-specific error code
<Description>
: brief description of the error
<Command>
: remote command causing the error
<Date>|<Time>
: date and time the error occurred
Usage:
Query only
SYSTem:ERRor[:NEXT]?
This command queries the most recent error queue entry and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
For details on error queues see
Chapter 10.1.7, "Status Reporting System"
Usage:
Query only
Configuring the R&S