![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 849](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985849.webp)
Remote Commands
R&S
®
FPL1000
849
User Manual 1178.3370.02 ─ 09
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<SearchLimit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
left diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation:
See
"Search Limits (Left / Right)"
See
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Limit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
right diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation:
See
"Search Limits (Left / Right)"
See
Setting Basic Spectrum RF Measurement Parameters