![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 841](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985841.webp)
Remote Commands
R&S
®
FPL1000
841
User Manual 1178.3370.02 ─ 09
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT2:LINK ON
Manual operation:
See
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>
<State>
This command links delta marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, delta marker <m1> changes its
horizontal position to the same value.
Suffix:
<n>
.
<ms>
source marker, see
<md>
destination marker, see
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT4:LINK:TO:MARK2 ON
Links the delta marker 4 to the marker 2.
Manual operation:
See
CALCulate<n>:DELTamarker<m>:MODE
<Mode>
This command defines whether the position of a delta marker is provided as an abso-
lute value or relative to a reference marker.
Note that when the position of a delta marker is
queried
, the result is always an abso-
lute value (see
Suffix:
<n>
.
<m>
irrelevant
Parameters:
<Mode>
ABSolute
Delta marker position in absolute terms.
RELative
Delta marker position in relation to a reference marker.
*RST:
RELative
Setting Basic Spectrum RF Measurement Parameters