![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 707](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985707.webp)
Remote Commands
R&S
®
FPL1000
707
User Manual 1178.3370.02 ─ 09
CALCulate<n>:STATistics:SCALe:X:RLEVel
<RefLevel>
This command sets the reference level for statistical measurements. The effects are
identical to
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
.
Note that in case of statistical measurements the reference level applies to the x-axis.
Suffix:
<n>
.
Parameters:
<RefLevel>
The unit is variable.
If a reference level offset is included, the range is adjusted by
that offset.
Range:
-130 dBm to 30 dBm
*RST:
0 dBm
Default unit: dBm
Example:
CALC:STAT:SCAL:X:RLEV -60dBm
Manual operation:
See
CALCulate<n>:STATistics:SCALe:Y:LOWer
<Magnitude>
This command defines the lower vertical limit of the diagram.
Suffix:
<n>
.
Parameters:
<Magnitude>
The number is a statistical value and therefore dimensionless.
Range:
1E-9 to 0.1
*RST:
1E-6
Example:
CALC:STAT:SCAL:Y:LOW 0.001
Manual operation:
See
CALCulate<n>:STATistics:SCALe:Y:UNIT
<Unit>
This command selects the unit of the y-axis.
Suffix:
<n>
.
Parameters:
<Unit>
PCT | ABS
*RST:
ABS
Example:
CALC:STAT:SCAL:Y:UNIT PCT
Sets the percentage scale.
Manual operation:
See
Configuring and Performing Measurements