The Spectrum Application (RF Measurements)
R&S
®
FPL1000
310
User Manual 1178.3370.02 ─ 09
attenuation values cause the inherent noise (i.e the noise floor) to rise and the sensitiv-
ity of the analyzer decreases.
The sensitivity of a signal analyzer is directly influenced by the selected RF attenua-
tion. The highest sensitivity is obtained at an RF attenuation of 0 dB. Each additional
10 dB step reduces the sensitivity by 10 dB, i.e. the displayed noise is increased by 10
dB. To measure a signal with an improved signal-to-noise ratio, decrease the RF
attenuation.
For ideal sinusoidal signals, the displayed signal level is independent of the RF attenu-
ation.
Depending on the type of measurement you must find a compromise between a low
noise floor and high intermodulation levels, and protecting the instrument from high
input levels. You achieve this best by letting the R&S
FPL1000 determine the optimum
7.5.1.3
Scaling
In a linear display, the measurement values are distributed linearly throughout the grid.
That means the entire range of measured values is divided by the number of rows in
the grid (10) and each row corresponds to 1/10 of the total range. Linear scaling is use-
ful to determine precise levels for a small range of values. However, if large and small
values appear in the same display, it is difficult to determine individual values precisely
or to distinguish values that are close together.
In a logarithmic display, smaller values are distributed among a much larger area of the
display, while large values are condensed to a smaller area. Now it is much easier to
distinguish several lower values, as they are spread over a wider area. Logarithmic
scaling is useful when large ranges of values must be combined in one display. Loga-
rithmic scaling is best applied to measurement values in logarithmic units (dB, dBm
etc.).
In addition to linear or logarithmic scaling, the vertical axis can be set to display either
absolute or relative values. Absolute values show the measured levels, while relative
values show the difference between the measured level and the defined reference
level. Relative values are indicated in percent for linear scaling, and in dB for logarith-
mic scaling.
7.5.2
Amplitude Settings
Access
: "Overview" > "Amplitude"
Amplitude settings determine how the R&S
FPL1000 must process or display the
expected input power levels.
The remote commands required to define these settings are described in
ter 11.8.3.1, "Amplitude Settings"
Amplitude and Vertical Axis Configuration