![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 199](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985199.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
199
User Manual 1178.3370.02 ─ 09
File contents
Explanation
y-Axis;LOG;
Level Range;100.000000;dB
Trace settings
Trace Mode;CLR/WRITE;
Detector;RMS;
Sweep Count;0;
Trace 1:;
x-Unit;Hz;
y-Unit;dBm;
List evaluation settings
Margin;200;
Peak List margin
Reference range settings
RefType; CPOWER;
Reference power type
TxBandwidth;3840000;;Hz
Channel power settings
Filter State; ON;
Alpha;0.22;
PeaksPerRange;1;
Max. number of peaks per range to be detected
Values;2;
Number of detected peaks
File data section
0;-12750000;-2515000;30000;13242367500;-43.844
722747802734;-0.33028793334960938;49.6697120
66650391;FAIL;
Measured peak values:
<range number>;
<start frequency>;
<stop frequency>;
<resolution bandwidth of range>;
<frequency of peak>;
<absolute power in dBm of peak>;
<relative power in dBc of peak>; (related to the
channel power)
<distance to the limit line in dB>; (positive value
means above the limit)
<limit fail (pass = 0, fail =1)>;
2;2515000;12750000;30000;13257632500;-43.8447
22747802734;-0.33028793334960938;49.66971206
6650391;FAIL;
7.2.8
Spurious Emissions Measurement
Spurious Emissions measurements monitor unwanted RF products outside the
assigned frequency band generated by an amplifier.
Measurements and Results