![Ronde & Schwarz 1304.0004.03 Скачать руководство пользователя страница 163](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985163.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
163
User Manual 1178.3370.02 ─ 09
Figure 7-27: SEM results for multiple sub blocks
Retrieving results via remote control
The measurement results of the spectrum emission mask
test
can be retrieved using
the
CALC:LIM:FAIL?
command from a remote computer; see
on page 914 for a detailed description.
The
power
result for the reference range can be queried using
CALC:MARK:FUNC:POW:RES? CPOW
;
The
peak
power for the reference range can be queried using
CALC:MARK:FUNC:POW:RES? PPOW
, see
CALCulate<n>:MARKer<m>:FUNCtion:
The measured power
trace
can be queried using
TRAC:DATA?
and
TRAC:DATA:X?
,
The measured
peak power list
can be queried using
TRAC:DATA? LIST
, see
7.2.7.4
SEM Basics
Some background knowledge on basic terms and principles used in SEM measure-
ments is provided here for a better understanding of the required configuration set-
tings.
................................................................................. 164
Limit Lines in SEM Measurements
........................................................................166
......................................................................................168
SEM with Multiple Sub Blocks ("Multi-SEM")
........................................................ 169
Measurements and Results