Matos
®
Fridge Installation & Guide
PLEASE READ
Unpacking & Setting Up
When installing the fridge, please ensure that the space is adequate with a minimum 5cm ventilation
gap either side of the machine and above.
Install the two rubber “bungs” in the port hole on the side of the fridge (both inside and outside) and
place the shelves evenly throughout the fridge to assist with air-flow.
If you have a Matos Cloud fridge, install the supplied fuse in the
3.15A battery fuse holder on the back of the fridge.
The use of a surge protector is required for warranty requirements.
You may require force to insert the plug into some models of surge
protector fully to ensure the plug is connected correctly.
This fridge has been pre-programmed and tested by Rollex Medical technicians before being
dispatched to you and no adjustment should be necessary. The refrigerator will not need
reprogramming after being unplugged or a power cut.
Typically the set point is set to 3.5°C and the refrigerator will then cycle above this point. It is normal
for the temperature on the display to change as it shows the current internal probe temperature.
Typically it will cycle between 3.5° – 6°C.
It is recommended you adjust the front feet or install furniture cups to level / tilt the cabinet so that
the door will close from approximately 1 cm.
NOTE: Placing the fridge on carpet is not recommended.
IMPORTANT: Overstocking, poor airflow and doors left ajar can lead to ice buildup and
this may require being manually defrosted.