59 / 341 - EN -
4.3.
Connecting the connection cable
Step
Description
1
Insert the plug of the control unit into the socket on the kiln.
2
The plug and socket have a geometric coding – the plug and socket only
fit together in one position.
3
The wide lug on the plug must be at the top in the 12 o’clock position to fit
into the wide recess on the plug, also in the 12 o’clock position.
4
You may have to twist the plug a little until it clicks completely into the
socket.
5
Tighten the outer screw ring on the plug clockwise.
4.4.
Control unit extension cable
•
If the bracket for the control unit is attached at the operation location (near the kiln or mounted on a wall), the
cable can be extended using an extension cable.
•
The extension cable for the control systems is optionally available in lengths of 2.5, 5 or a maximum of 10
meters.
•
Note on extension cables and environmental influences due to EMC:
To meet the requirements for electromagnetic compatibility (EMC), the length of the controller
connection cable should not exceed 3 meters.
If the controller is connected to the kiln with an extension cable, it should be ensured that there is no
electrical device in the direct vicinity of the cable (electromagnetic stray field).
Otherwise, a loss of accuracy of up to 3°C could occur.
4.5.
Third-party kiln note
The wiring of the mating CPC 14 or CPC 19 socket on the kiln can vary between kiln manufacturers!
Attention!
If the assignments of the controller and kiln do not match, then there is a risk of damage to both the controller and kiln.
5.
Installation
5.1.
Switching the control unit on & off
The rocker switch for switching the control unit on and off is located on the underside of the housing.
Switching the
control unit on
Set the rocker switch to position “I”.
Switching the
control unit off
Set the rocker switch to position “0”.