Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
204
User Manual 1178.5989.02 ─ 06
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Return values:
<PeakPosition>
Position of the peaks on the y-axis. The unit depends on the
measurement.
Usage:
Query only
MMEMory:STORe<n>:PEAK
<FileName>
This command exports the marker peak list to a file.
Suffix:
<n>
.
Parameters:
<FileName>
String containing the path,name and extension of the target file.
Example:
MMEM:STOR:PEAK 'test.dat'
Saves the current marker peak list in the file
test.dat
.
Manual operation:
See
9.8
Retrieving Results
The following commands can be used to retrieve the results of the I/Q Analyzer mea-
surement.
Storing large amounts of I/Q data
When storing large amounts of I/Q data to a file, consider the following tips to improve
performance:
●
If capturing and storing the I/Q data is the main goal of the measurement and eval-
uation functions are not required, use the basic I/Q data acquisition mode (see
●
Use a HiSlip or raw socket connection to export the data from the R&S
ZNL to a
PC.
●
Export the data in binary format rather than ASCII format (see
for Returned Values: ASCII Format and Binary Format"
●
Use the "Compatible" or "IQPair" data mode (see
Description for I/Q Data Files"
●
If only an extract of the available data is relevant, use the
command to store only the required section of data.
Retrieving Results