Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
157
User Manual 1178.5989.02 ─ 06
In case of max hold, min hold or average trace mode, you can set the number of single
measurements with
. Note that synchronization to the end of
the measurement is possible only in single sweep mode.
Suffix:
<n>
.
<t>
Parameters:
<Mode>
WRITe
Overwrite mode: the trace is overwritten by each sweep. This is
the default setting.
AVERage
The average is formed over several sweeps. The "Sweep/Aver-
age Count" determines the number of averaging procedures.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S
ZNL saves the sweep result in the trace mem-
ory only if the new value is greater than the previous one.
MINHold
The minimum value is determined from several measurements
and displayed. The R&S
ZNL saves the sweep result in the trace
memory only if the new value is lower than the previous one.
VIEW
The current contents of the trace memory are frozen and dis-
played.
BLANk
Hides the selected trace.
*RST:
Trace 1: WRITe, Trace 2-6: BLANk
Example:
INIT:CONT OFF
Switching to single sweep mode.
SWE:COUN 16
Sets the number of measurements to 16.
DISP:TRAC3:MODE WRIT
Selects clear/write mode for trace 3.
INIT;*WAI
Starts the measurement and waits for the end of the measure-
ment.
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE:HCONtinuous
<State>
This command turns an automatic reset of a trace on and off after a parameter has
changed.
The reset works for trace modes min hold, max hold and average.
I/Q Analysis