The skew measurement can only be performed if the following conditions are
met for both the active trace and the reference trace:
●
REAL
)
●
Time Domain is enabled (
CALCulate<Chn>:TRANsform:TIME:STATe
)
●
the Low Pass Step time domain transform is used
CALCulate<Chn>:TRANsform:TIME:
)
The latter, in turn, requires the stimulus grid to be harmonic. This can be achieved, for
example, using
ON
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON – Skew measurement enabled
OFF – Skew measurement disabled
*RST:
OFF
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:DTIMe:TARGet
<SkewRefTrace>
Selects the reference trace for the skew measurement.
See
for conditions on both the active and the refer-
ence trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SkewRefTrace>
Reference trace, identified by its name (within channel <Chn>).
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
7.3.2.5
CALCulate:EYE...
Defines the properties and retrieves the results of the eye diagram measurement provi-
ded with the
ZNB/ZNBT-K20.
................................................................................. 695
CALCulate<Chn>:EYE:EMPHasis:CURSor:POST<1|2>
.....................................................695
CALCulate<Chn>:EYE:EMPHasis:CURSor:PRE
............................................................... 695
CALCulate<Chn>:EYE:EMPHasis:STATe
........................................................................ 696
CALCulate<Chn>:EYE:EQUalization:CTLE:DC
.................................................................696
CALCulate<Chn>:EYE:EQUalization:CTLE:POLE<1|2>
.....................................................696
CALCulate<Chn>:EYE:EQUalization:CTLE:ZERO
.............................................................697
CALCulate<Chn>:EYE:EQUalization:STATe
.....................................................................697
SCPI Command Reference