GUI Reference
R&S
®
ZNB/ZNBT
517
User Manual 1173.9163.02 ─ 38
●
Limit check and display of the limit lines are independent of each other: With dis-
abled limit check, the limit line can still be displayed. If no limit lines are defined for
the active trace, the limit check can be switched on but will always PASS the trace.
●
Limit lines in "dB Mag" format are also valid for polar diagrams: if the correspond-
ing limit check is enabled it will show the correct PASS/FAIL state and limit fail
symbols (if any).
Figure 5-46: dB Mag limit line check in polar diagrams
If enabled, the
Remote command:
CALCulate<Chn>:LIMit:STATe
CALCulate<Chn>:LIMit:LOWer:STATe
CALCulate<Chn>:LIMit:UPPer:STATe
CALCulate<Chn>:LIMit:FAIL?
CALCulate:LIMit:FAIL:ALL?
CALCulate<Chn>:LIMit:STATe:AREA
Limit Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded limit. No fail beep can be gen-
erated if the limit check is switched off.
Remote command:
CALCulate<Chn>:LIMit:SOUNd[:STATe]
Clear Test
Resets the limit check results.
Remote command:
Global Check
Activates or deactivates the global limit check including upper/lower limits and ripple
limits. The global limit check is a composite limit check over all traces of the current
recall set. The result of the global check appears in a popup box whenever "Global
Check" is selected.
Lines Softtool