GUI Reference
R&S
®
ZNB/ZNBT
434
User Manual 1173.9163.02 ─ 38
Figure 5-42: Stimulus > Time Domain X-Axis softtool tab
left = Time representation
right = Distance representation
Time Start / Time Stop / Time Center / Time Span
Defines the display range for the time domain trace in time representation (see
●
"Time Start" is the lowest displayed time and corresponds to the left edge of the
Cartesian diagram.
●
"Time Stop" is the highest displayed time and corresponds to the right edge of the
Cartesian diagram.
●
"Time Center" corresponds to the center of the Cartesian diagram, i.e. ("Time
Start" + "Time Stop")/2.
●
"Time Span" corresponds to the diagram width, i.e. ("Time Stop" – "Time Start").
"Time Start" and "Time Stop" or "Time Center" and "Time Span" are alternative set-
tings.
Remote command:
CALCulate<Chn>:TRANsform:TIME:STARt
CALCulate<Chn>:TRANsform:TIME:STOP
CALCulate<Chn>:TRANsform:TIME:CENTer
CALCulate<Chn>:TRANsform:TIME:SPAN
Distance Start / Distance Stop / Distance Center / Distance Span
Defines the display range for the time domain trace in distance representation (see
"Distance Start" and "Distance Stop" or "Distance Center" and "Distance Span" are
alternative settings.
Stimulus Softtool