GUI Reference
R&S
®
ZNB/ZNBT
415
User Manual 1173.9163.02 ─ 38
Figure 5-38: Fixture Compensation dialog (for >4 ports)
Ports
Selects the ports for whom fixture compensation data shall be acquired.
Auto Length / Auto Length and Loss
"Auto Length" or "Auto Length and Loss" implies that a global electrical length offset
and loss is determined in analogy to the general offset compensation (see
Chapter 4.6.1.4, "Auto Length and Loss"
Remote command:
[SENSe:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe]
Direct Compensation
With "Direct Compensation", a frequency-dependent transmission factor is calculated;
see
"Auto Length and Loss vs. Direct Compensation"
Remote command:
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe]
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Prompt for Each Port
Determines how the R&S
ZNB/ZNBT performs the sweeps for a given termination type
●
If unchecked, it performs the sweeps for Open/Short without interruption, implicitly
assuming that
all
ports are terminated accordingly
●
If checked, it interrupts the data acquisition process after each port, which allows
you to modify the test setup (e.g. terminate the next measured port).
All Compensation On / All Compensation Off
Activates/deactivates the length and loss compensation for all ports.
"Active" column to activate/deactivate the length and
loss compensation for selected ports.
Offset Embed Softtool