GUI Reference
R&S
®
ZNB/ZNBT
353
User Manual 1173.9163.02 ─ 38
In this dialog enter the following parameters (according to the device to be controlled):
●
"SA": a slave address between 0 and 15 as 1 hex digit
●
"CMD": a command number between 0 and 255 as 2 hex digits
●
"DATA": the data part, 0 to 17 bytes (hex digit
pairs
)
●
"Expected Result Read Count of Bytes (decimal)": The number of bytes to be read
back after the command was sent (max. 16). If set to 0 the command is a pure
write command.
In Sweep Sequencer mode this section is hidden.
For details and background information see the "MIPI Alliance Specification for RF
Front-End Control Interface".
Remote command:
CONTrol<Ch>:RFFE<Bus>:COMMand:DATA
SEND
Use the "SEND" button to send the previously defined command, e.g. before starting
the sweep for the related channel.
On a R&S
ZN-B15/Z15
var. 03
, before the command is executed the related shunt
resistance is set to its minimum possible value.
Remote command:
CONTrol<Ch>:RFFE<Bus>:COMMand:SEND
(write-only) or
CONTrol<Ch>:RFFE<Bus>:COMMand:SEND?
<BytesToRead>
(with read-back)
RFFE Interface Voltage and Current Measurements
Defines the voltage and current measurements on the RFFE pins.
Channel Config Softtool