GUI Reference
R&S
®
ZNB/ZNBT
338
User Manual 1173.9163.02 ─ 38
Get Trace...
Opens a selection box containing all traces in the active recall set. The "dB Mag" val-
ues of the selected trace are used to define the transmission coefficients. Notice that if
you combine different channels with different sweep points, the analyzer may have to
interpolate or extrapolate the transmission coefficients.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:FEED
Import File...
Imports the transmission coefficients from a trace file. The imported file must be either
in Touchstone (
*.s<n>p
) or in
*.csv
Remote command:
MMEMory:LOAD:CORRection:TCOefficient
Recall... / Save...
You can save the displayed power loss list to a power meter correction list file with
extension (
*.pmcl
) and re-load it in later sessions.
Remote command:
MMEMory:LOAD:CORRection:TCOefficient
MMEMory:STORe:CORRection:TCOefficient
5.4.4 Use Cal Tab
Provides access to functions for (de-)activating and managing calibrations.
Background information
Refer to the following sections:
●
Cal Softtool