Command Reference
R&S
®
ZNB/ZNBT
1207
User Manual 1173.9163.02 ─ 38
Query parameters:
<Response>
CH1Data | CH2Data | CH3Data | CH4Data | CH1Mem |
CH2Mem | CH3Mem | CH4Mem | MDATa1 | MDATa2 |
MDATa3 | MDATa4 | MDATa5 | MDATa6 | MDATa7 | MDATa8
Response data of the selected trace, see
.
The data is transferred in the data format defined via
. The unit is the default
unit of the measured parameter; see
Example:
SWE:POIN 20
Create a trace with 20 sweep points, making the created trace
the active trace of channel 1 (omitted optional mnemonic
SENSe1
).
CALC:FORM MLIN; :FORM ASCII; FORM:DEXP:SOUR
FDAT
Select the trace data format: linear magnitude values, ASCII for-
mat and formatted trace data (1 value per sweep point).
TRAC? CH1DATA
Query the 20 response values of the created trace according to
the previous format settings.
Usage:
Query only
TRACe[:DATA]:STIMulus[:ALL]?
<Stimulus>
Returns the stimulus values of the active data trace or memory trace (see
Note:
To read the stimulus values of an arbitrary data or memory trace, use
Query parameters:
<Stimulus>
CH1Data | CH2Data | CH3Data | CH4Data | CH1Mem |
CH2Mem | CH3Mem | CH4Mem | MDATa1 | MDATa2 |
MDATa3 | MDATa4 | MDATa5 | MDATa6 | MDATa7 | MDATa8
Stimulus data of the selected trace, see
.
The data is transferred in the data format defined via
Example:
SWE:POIN 20
Create a trace with 20 sweep points, making the created trace
the active trace of channel 1 (omitted optional mnemonic
SENSe1
).
TRAC:STIM? CH1DATA
Query the 20 stimulus values of the created trace. In the default
format setting, the data is returned as a comma-separated list of
10-digit ASCII values.
Usage:
Query only