Command Reference
R&S
®
ZNB/ZNBT
1120
User Manual 1173.9163.02 ─ 38
Parameters:
<RecCalPower>
Reference receiver cal power
Range:
Depending on source power range of the analyzer
and the power range of the external power meter.
*RST:
0 dBm
Default unit: dBm
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation:
See
"Reference Receiver Cal Power"
SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect
<CalPower>
Qualifies how to define the source power which the R&S
ZNB/ZNBT uses to perform
the first calibration sweep of the source power calibration ("Reference Receiver Cal
Power").
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
Parameters:
<CalPower>
CPOWer | PPOWer
PPOWer
–
use cal power value defined via
POWer<PhyPt>:CORRection:PPOWer
CPOWer
–
use port power value
*RST:
PPOWer
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation:
See
"Reference Receiver Cal Power"
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for port number
<PhyPt>
. The setting command is disabled unless a source power calibration for the
analyzer port has been performed (
SOURce<Ch>:POWer<PhyPt>:CORRection[:
). The query always returns a result.
To enable or disable a source power calibration for an external generator use
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe]
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
SCPI Command Reference