SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:COUNt?
.................................. 1122
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DEFine<ListNo>
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DELete<ListNo>:ALL
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DELete<ListNo>[:DUMMy]
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:FEED
...................................... 1124
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:INSert<ListNo>
SOURce:POWer:CORRection:PMETer:ID
<PowerMeter>
Selects an external power meter for the scalar source power calibration (flatness cali-
bration).
Note:
The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Parameters:
<PowerMeter>
Number of external power meter. The parameters UP, DOWN,
MIN, MAX are not available for this command.
Range:
1 to number of configured external generators
*RST:
The power meter selection is not changed by a
reset of the analyzer.
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:
Manual operation:
See
SOURce<Ch>:POWer:CORRection:DATA
<Source>, <CorrData>...
SOURce<Ch>:POWer:CORRection:DATA:PORT<PhyPt>
<Source>, <CorrData>...
Reads or writes scalar source power correction data sets. A power correction data set
contains n real values where:
●
Each value corresponds to the ratio of the actual power at the reference plane
(value provided by the used source) to the uncalibrated power in dB.
●
The number n is equal to the number of sweep points.
Increasing (decreasing) the values in the correction data sets increases (decreases)
the power at the reference plane. Writing correction data (the setting command) fails if
the number of transferred values is not equal to the number of sweep points.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Physical port number, used to select a specific frequency axis in
arbitrary mode (with option R&S
ZNB/ZNBT-K4).
Parameters:
<Source>
Identifier for the source of the calibrated wave:
'
A<n>
' or '
PORT<n>
' denote correction data for the analyzer port
no.
<n>
.
'
ESRC<n>
' or '
GEN<n>
' denote correction data for the external
generator no.
<n>
.
SCPI Command Reference