Installation and Commissioning
R&S
®
TS7124AS
28
Instructions Handbook 1179.2948.02 ─ 01
If you remove the chamber from its location, follow the instructions in
6.3.2
Rack Mounting
With the optional rack mounting kit R&S
TS-F24-Z1 (order no. 1526.6942.02), you can
mount the chamber in a standard 19'' rack.
In the rack, the kit requires a height of 10
HU (17.5'' or 444.5
mm). It consists of a
metal rack cover sheet and a set of washers and torx screws. The metal sheet has a
shape that goes around the chamber's front door. The sheet fits with the threaded
holes in the front of the chamber's body and with threaded holes in the frame of a stan-
dard rack.
Figure 6-3: Chamber fixed in a 19'' rack by the rack mounting kit
Red arrows = Metal sheet of the rack mounting kit
Inside the rack, the chamber must rest on stable rails that jointly can carry a load of at
least 40
kg. Fix the chamber to these rails (or to the rack) with screws and brackets
As a
prerequisite
for installation, the 19'' rack with universal square holes must be
equipped with
M5 cage nuts
:
Mounting the Chamber