Remote control commands
R&S
®
SMM100A
968
User Manual 1179.1341.02 ─ 05
:SYSTem:NINFormation?
Queries the oldest information message ("Error History > Level > Info") in the error/
event queue.
Return values:
<NextInfo>
string
Example:
:SYSTem:NINFormation?
Queries the oldest entry in the info message queue.
Response: 90,"Info;=== Instrument startup...
==="
Information message containing error number 90, that states,
that the instrument startup is complete.
Usage:
Query only
:SYSTem:ULOCk
<Mode>
Locks or unlocks the user interface of the instrument.
Parameters:
<Mode>
ENABled | DONLy | DISabled | TOFF | VNConly
ENABled
Unlocks the display, the touchscreen and all controls for the
manual operation.
DONLy
Locks the touchscreen and controls for the manual operation of
the instrument. The display shows the current settings.
VNConly
Locks the touchscreen and controls for the manual operation,
and enables remote operation over VNC. The display shows the
current settings.
TOFF
Locks the touchscreen for the manual operation of the instru-
ment. The display shows the current settings.
DISabled
Locks the display, the touchscreen and all controls for the man-
ual operation.
*RST:
n.a. (factory preset: ENABled)
Example:
:SYST:ULOC DIS
Activates the user interface lock, including display and controls.
Manual operation:
See
See
"Enabling a locked user interface for manual operation"
SYSTem subsystem