Remote control commands
R&S
®
SMM100A
940
User Manual 1179.1341.02 ─ 05
Manual operation:
See
[:SOURce<hw>]:POWer:MODE
<Mode>
Selects the operating mode of the instrument to set the output level.
Parameters:
<Mode>
CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW
and
FIXed
are synonyms.
To set the output level value, use the command
SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude]
SWEep
Sets sweep mode.
Set the range and current level with the commands:
*RST:
CW
Example:
See
Example "Setup an RF frequency or power sweep"
Manual operation:
See
See
"State (RF frequency / level sweep)"
[:SOURce<hw>]:POWer:PEP?
Queries the PEP "
P
eak
E
nvelope
P
ower) of digital modulation or digital standards at
the RF output. This value corresponds to the level specification, displayed in the status
bar (header).
Return values:
<Pep>
float
Example:
POW:PEP?
Response:
"4"
The PEP value of digital modulation is 4 dBm at the RF output.
Usage:
Query only
[:SOURce<hw>]:POWer:POWer
<Power>
Sets the level
at the RF output
connector.
This value does not consider a specified offset.
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
sets the level of the "Level" display, that means the level containing offset.
SOURce subsystem