![Rohde & Schwarz SMM100A Скачать руководство пользователя страница 870](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939870.webp)
Remote control commands
R&S
®
SMM100A
870
User Manual 1179.1341.02 ─ 05
Table 13-4: Value range
Impairments
Min [dB]
Max [dB]
Resolution
Digital
-30
30
0.01
Analog
-10
10
0.01
Parameters:
<Angle>
float
Range:
-30 to 30
Increment: 0.01
*RST:
0
Default unit: DEG
Example:
SOURce:BB:IMPairment:RF1:QUADrature:ANGLe -5
Sets the quadrature offset to -5 degrees.
Manual operation:
See
[:SOURce]:BB:IMPairment:BBMM<ch>:SKEW
<Skew>
[:SOURce]:BB:IMPairment:IQOutput<ch>:SKEW
<Skew>
[:SOURce]:BB:IMPairment:RF<ch>:SKEW
<Skew>
Sets a delay between the Q vector and the I vector of the corresponding stream.
Parameters:
<Skew>
float
Range:
-500E-9 to 500E-9
Increment: 1E-12
*RST:
0
Manual operation:
See
[:SOURce<hw>]:IQ:IMPairment[:STATe]
<State>
[:SOURce]:BB:IMPairment:BBMM<ch>:STATe
<State>
[:SOURce]:BB:IMPairment:IQOutput<ch>:STATe
<State>
[:SOURce]:BB:IMPairment:RF<ch>:STATe
<State>
Activates the impairment or correction values
LEAKage
,
QUADrature
and
IQRatio
for the corresponding stream.
Parameters:
<State>
1 | ON | 0 | OFF
*RST:
0
Example:
SOURce1:BB:IMPairment:RF1:STATe ON
Activates digital impairment.
Manual operation:
See
SOURce subsystem