Remote control commands
R&S
®
SMM100A
676
User Manual 1179.1341.02 ─ 05
:CALibration<hw>:DEBug
<State>
Activates logging of the internal adjustments.
Suffix:
<hw>
.
[1]
Optional suffix
Setting parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
CALibration:DEBug 1
Usage:
Setting only
Manual operation:
See
:CALibration<hw>:IQModulator:LOCal?
Starts adjustment of the I/Q modulator for the currently set frequency and baseband
gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q imbalance and
quadrature.
Return values:
<Local>
1 | ON | 0 | OFF
*RST:
0
Example:
// Start adjustment for the I/Q modulator
// for the currently set frequency
CALibration:IQModulator:LOCal?
// 0
// Adjustments are performed successfully
Usage:
Query only
Manual operation:
See
"Adjust I/Q Modulator Current Frequency"
:CALibration<hw>:FREQuency:CONVerter:EXTernal?
Queries the calibration state of the connected external instrument.
External instruments can be an I/Q upconverter R&S
SZU or an external frontend, e.g.,
R&S
FE44S.
Return values:
<Success>
1 | ON | 0 | OFF
*RST:
0
CALibration subsystem