Network operation and remote control
R&S
®
SMM100A
612
User Manual 1179.1341.02 ─ 05
LAN Status
The LED indicates the LXI status.
"green"
Normal operation
"green (flashing)"
Device identification
"red"
LAN fault
Hostname / MAC Address / IP Address
See
LAN Reset
Initiates the network configuration reset mechanism for the instrument and resets the
hostname, MAC address, and IP address.
According to the LXI standard, a LAN reset must set the following network parameters
to a default state:
Parameter
Value
TCP/IP mode
DHCP + Auto IP address
Dynamic DNS
Enabled
ICMP ping
Enabled
Password for LAN configuration
LxiWebIfc
The LAN reset also resets the following parameters for the vector signal generator:
Parameter
Hostname
Instrument-specific hostname
Description
Vector signal generator
Negotiation
Auto detect
VXI-11 discovery
Enabled
12.6.2
LXI browser settings
The LXI browser functionality enables you to control the instrument remotely over a
web browser.
Access:
► In the address field of the browser, enter the URL address, composed of the instru-
ment's IP address or host name, for example
http://10.113.1.151
.
Note:
Do not add the missing zeros in the IP address, while opening the instru-
ment homepage.
LXI settings