Test sequence
R&S®SML / R&S®SMV03
1147.7621.12 10.42
E-6
Nonharmonic suppression
Test setup
Test setup 1 with spectrum analyzer (see Performance test R&S
SML03 / R&S SMV03)
Settings on R&S SMV03
- UTILITIES REF OSC SOURCE EXTERNAL
- LEVEL 10 dBm
- VECTORMOD STATE IQCW
- FREQ
Test frequecy of nonharmonics
Test frequencies of
nonharmonics: 250.0MHz, 250.0000001MHz
Setting on spectrum analyzer
- Reference level = test level + 3 dB, 10 dB/div
- Test frequency = 200 MHz, 800 MHz, 1600 MHz, 2400 MHz, 3200
MHz
- Span = 1 MHz
- Switch on average: 5 samples
Measurement
See Performance test R&S SML03 / R&S SMV03.
Evaluation
See Performance test R&S SML03 / R&S SMV03.
Broadband noise
Test setup
Test setup 3 (see Performance test R&S SML03 / R&S SMV03)
Settings on
R&S SMV03
-
UTILITIES REF OSC SOURCE EXTERNAL
-
VECTORMOD STATE IQCW
-
LEVEL 0 dBm (or . level according to mixer specification)
-
FREQUENCY 1 GHz (or any test frequency)
Test method
See Performance test R&S SML03 / R&S SMV03.
Measurement
See Performance test R&S SML03 / R&S SMV03
Evaluation
See Performance test R&S SML03 / R&S SMV03.
Содержание SML01
Страница 4: ......
Страница 14: ......
Страница 22: ......
Страница 32: ......
Страница 34: ......
Страница 36: ......
Страница 38: ......
Страница 40: ......
Страница 42: ......
Страница 44: ......
Страница 56: ......
Страница 62: ......
Страница 77: ...R S SML R S SMV03 Menu Summary for R S SML 1090 3123 12 E 6 3 15 Menu Summary for R S SML ...
Страница 78: ...Menu Summary for R S SMV03 R S SML R S SMV03 1090 3123 12 E 6 3 16 Menu Summary for R S SMV03 ...
Страница 148: ......
Страница 240: ......
Страница 242: ......