Remote Control Commands
R&S
®
SMF100A
583
Operating Manual 1167.2319.02 ─ 12
Return values:
<Resource>
string
Example:
SYSTem:COMMunicate:HISLip:RESource?
Response:
"TCPIP::192.1.2.3::hislip0::INSTR"
Usage:
Query only
Manual operation:
See
:SYSTem:COMMunicate:USB:RESource?
Queries the visa resource string for remote control via the USB interface.
Return values:
<Resource>
string
Example:
SYSTem:COMMunicate:USB:RESource?
queries the VISA resource string for remote control via the USB
interface.
Response:
"USB::72::000000::INSTR"
Usage:
Query only
Manual operation:
See
:SYSTem:COMMunicate:SERial:RESource?
Queries the visa resource string for the serial remote control interface. This string is
used for remote control of the instrument.
Return values:
<Resource>
string
Example:
SYSTem:COMMunicate:SERial:RESource?
queries the VISA resource string.
Response:
"ASRL1::INSTR"
Usage:
Query only
Manual operation:
See
:SYSTem:COMMunicate:SERial:BAUD
<Baud>
Sets the baudrate for the serial remote control interface.
Parameters:
<Baud>
2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200
*RST:
115200
Example:
SYSTem:COMMunicate:SERial:BAUD 115200
determines 115200 baudrate.
Manual operation:
See
SYSTem Subsystem