![Rohde & Schwarz SMF-B1 Скачать руководство пользователя страница 591](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740591.webp)
Remote Control Commands
R&S
®
SMF100A
576
Operating Manual 1167.2319.02 ─ 12
Return values:
<All>
string
0
"No error", i.e. the error queue is empty
positive value
Positive error numbers denote device-specific errors
negative value
Negative error numbers denote error messages defined by
SCPI.
Example:
SYST:ERR:CODE:ALL
queries all entries in the error queue.
Response: 0
no errors have occurred since the error queue was last read out.
Usage:
Query only
:SYSTem:ERRor:CODE[:NEXT]?
Queries the oldest entry in the error queue and then deletes it. Only the error number
is returned and not the entire error text.
Return values:
<Next>
string
0
"No error", i.e. the error queue is empty
positive value
Positive error numbers denote device-specific errors
negative value
Negative error numbers denote error messages defined by
SCPI.
Example:
SYST:ERR:CODE
queries the oldest entry in the error queue.
Response: 0
No errors have occurred since the error queue was last read out.
Usage:
Query only
:SYSTem:ERRor:COUNt?
Queries the number of entries in the error queue. If the error queue is empty, '0' is
returned.
Return values:
<Count>
string
Example:
SYST:ERR:COUN
queries the number of entries in the error queue.
Response: 1
One error has occurred since the error queue was last read out.
SYSTem Subsystem