Remote Control Commands
R&S
®
SMF100A
555
Operating Manual 1167.2319.02 ─ 12
Example:
ROSC:OUTP:SOUR EXT
the reference oscillator signal is output at the REF OUT connec-
tor.
Manual operation:
See
7.13.18 SOURce:SWEep Subsystem
The
SOURce:
subsystem contains the commands for configuring RF sweep signals.
●
The keyword
[:FREQuency]
can be ommitted, then the commands are SCPI-
compliant.
●
To activate a RF sweep mode, use the following commands:
–
RF frequency sweep:
SOURce:FREQuency:MODE SWEep
(
SOURce:FREQuency:MODE CW
(off))
–
RF level sweep:
SOURce:POWer:MODE SWEep
(
SOURce:POWer:MODE CW
(off))
●
All sweeps, including the LF sweep, can be set independently from each other.
This example shows how to set up a frequency sweep.
1. Set the sweep range.
[SOURce:]FREQuency:CENTer 200 MHz
[SOURce:]FREQuency:SPAN 300 MHz
2. Select linear or logarithmic spacing.
[SOURce:]SWEep[:FREQuency]:SPACing LIN
3. Set the step width and dwell time.
[SOURce:]SWEep[:FREQuency]:STEP:LINear 20 MHz
[SOURce:]SWEep[:FREQuency]:DWELl 12 ms
4. Select the trigger mode.
TRIGger:]FSWeep:SOURce SINGle
5. Select the sweep mode and activate the sweep.
[SOURce:]SWEep[:FREQuency]:MODE AUTO
[SOURce:]FREQuency:MODE SWEep
6. Trigger the sweep.
[SOURce:]SWEep[:FREQuency]:EXECute
It is recommended that you switch off the "Start/Stop Display Update" for optimum
sweep performance, especially with short dwell times (
SYSTem:DISPlay:UPDate
OFF
).
SOURce Subsystem