Remote Control Commands
R&S
®
SMF100A
461
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Type>
NRP | NRVD | HP437 | ML2438 | P44X
*RST:
NRP
Example:
SOUR:CORR:PMET:TYPE NRVD
the specified power meter is used for user correction.
Manual operation:
See
" User Correction Value - User Correction "
[:SOURce<hw>]:CORRection[:STATe]
<State>
Activates/deactivates level correction. Level correction is performed using the table
which has been selected with the command
[:SOURce<hw>]:CORRection:CSET[:
.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SOUR:CORR:CSET '/var/user/ucor1'
selects the table
ucor1
.
SOUR:CORR ON
activates user correction.
Manual operation:
See
[:SOURce<hw>]:CORRection:VALue?
Queries the current value for user correction.
Return values:
<Value>
float
Range:
-100 to 100
Increment: 0.01
*RST:
0
Example:
CORR:VAL?
queries the value currently used for level correction.
Response:
-3
the correction value is - 3 dB.
Usage:
Query only
Manual operation:
See
" User Correction Value - User Correction "
[:SOURce<hw>]:CORRection:ZERoing:STATe
<State>
Activates the zeroing procedure before filling the user correction data acquired by a
sensor.
Parameters:
<State>
0 | 1 | OFF | ON
SOURce Subsystem