Remote Control Commands
R&S
®
SMF100A
423
Operating Manual 1167.2319.02 ─ 12
Example:
SENS1:SWE:TIM:PULS:THR:REF 40
sets the medial reference level to 40% of the overall pulse level.
Options:
Option R&S SMF-K28
Manual operation:
See
" Mesial - Pulse Data Analysis "
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:POWer:
LREFerence
<LReference>
Sets the lower reference level in terms of percentage of the overall pulse level. The
proximal power defines the start of the rising edge and the end of the falling edge of
the pulse.
Note:
This parameter is only avalaible in time measurement mode and R&S NRP-Z81
power sensors.
Parameters:
<LReference>
float
Range:
0.0 to 100.0
Increment: 0.01
*RST:
10.0
Example:
SENS:SWE:TIM:PULS:THR:LREF 10
sets the lower reference level to 10%.
Options:
Option R&S SMF-K28
Manual operation:
See
" Proximal - Pulse Data Analysis "
:SENSe[:POWer]:SWEep:POWer:SPACing[:MODE]
<Mode>
Queries the sweep spacing for the power versus power measurement. The setting is
fixed to LINear.
Parameters:
<Mode>
LINear
*RST:
LINear
Example:
SENS:SWE:POW:SPAC?
Options:
Option R&S SMF-K28
Manual operation:
See
:SENSe[:POWer]:SWEep:POWer:STARt
<Start>
Sets the start level for the power versus power measurement.
Parameters:
<Start>
float
Range:
-145 to 20
Increment: 0.01
*RST:
-40 dBm
Power Sensor Measurement Subsystems