Remote Control Commands
R&S
®
SMF100A
383
Operating Manual 1167.2319.02 ─ 12
Example:
OUTP:ALL OFF
switches off all RF output signals.
:OUTPut<hw>:AFIXed:RANGe:LOWer?
Queries the minimum level which can be set when the attenuator is fixed, see
.
Return values:
<Lower>
float
Increment: 0.01
Example:
OUTP:AFIX:RANG:LOW?
queries the minimum level for the FIXed setting.
Example:
Response: -50
The minimum level is -50 dBm.
Usage:
Query only
Manual operation:
See
" Fixed Range in (PEP): - Attenuation Settings "
:OUTPut<hw>:AFIXed:RANGe:UPPer?
Queries the maximum level which can be set when the attenuator is fixed, see
.
Return values:
<Upper>
float
Increment: 0.01
Example:
OUTP:AFIX:RANG:UPP?
queries the maximum level for the FIXed setting for the RF out-
put.
Example:
Response: -27
The maximum level is -27 dBm.
Usage:
Query only
Manual operation:
See
" Fixed Range in (PEP): - Attenuation Settings "
:OUTPut<hw>:ALC:SEARch:MODE
<Mode>
Activates/deactivates the RF output during the power search.
Parameters:
<Mode>
NORMal | MINimum
*RST:
NORMal
OUTPut Subsystem