Remote Control Commands
R&S
®
SMF100A
378
Operating Manual 1167.2319.02 ─ 12
<DestinationFile>
string
String containing the path and name of the target file. The path
can be relative or absolute.
If
<DestinationFile>
is not specified, the
<SourceFile>
is
copied to the current directory, queried with the
Note:
Existing files with the same name in the destination direc-
tory are overwritten without an error message.
Example:
See
"Working with Files and Directories"
Usage:
Setting only
SCPI confirmed
Manual operation:
See
:MMEMory:DATA
<Filename>, <BinaryBlock>
:MMEMory:DATA?
<Filename>
The setting command writes the block data
<BinaryBlock>
to the file identified by
<Filename>
.
Set the GPIB-bus terminator to
EOI
to ensure correct data transfer.
The query command transfers the specified file from the instrument to the GPIB-bus
and then on to the controller. It is important to ensure that the intermediate memory on
the controller is large enough to take the file. The setting for the GPIB-bus terminator is
irrelevant.
Tip:
Use this command to read/transfer stored instrument settings or waveforms
directly from/to the instrument.
Parameters:
<BinaryBlock>
#<number><length_entry><data>
#
: Hash sign; always comes first in the binary block
<number>
: the first digit indicates how many digits the subse-
quent length entry has
<length_entry>
: indicates the number of subsequent bytes
<data>
: binary block data for the specified length.
For files with a size with more than nine digits (gigabytes), the
instrument allows the syntax
#(<Length>)
, where
<Length>
is the file size in decimal format.
Parameters for setting and query:
<Filename>
string
String parameter to specify the name of the file.
MMEMory Subsystem