![Rohde & Schwarz SMF-B1 Скачать руководство пользователя страница 286](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740286.webp)
Instrument Function
R&S
®
SMF100A
271
Operating Manual 1167.2319.02 ─ 12
All other analog modulations are deactivated during chirp modulation
When chirp modulation is activated, any active analog modulation is automatically
switched off.
The Sample&Hold measurement for the ALC is performed at the chirp center fre-
quency. Therefore, the frequency response of the RF path can result in level errors for
the FM modulated carrier especially with high chirp bandwidths.
5.5.7.1
Chirp Modulation Settings
Access:
► Select "Modulation > config... > Chirp Modulation".
The "Chirp Modulation" dialog contains the parameters for configuring the modula-
tion and trigger settings.
State - Chirp Modulation
Activates/deactivates chirp modulation
Note:
Any active modulation is automatically switched off when chirp modulation is
activated.
The pulse generator signal is provided at the PULSE OUT output, the sync signal at
the SYNC output.
These outputs are automatically switched on/off according to the chirp modulation
state. The FM modulation signal can be output at the LF connector.
Remote command:
Direction - Chirp Modulation
Selects the direction of the chirp modulation.
Modulation