![Rohde & Schwarz SMF-B1 Скачать руководство пользователя страница 267](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740267.webp)
Instrument Function
R&S
®
SMF100A
252
Operating Manual 1167.2319.02 ─ 12
All level correction values for the given frequency values are measured using the
Power Sensor and automatically filled in the selected list after the "Execute" button is
pressed. The list is automatically stored and recalled again after filling.
Fill User Correction Data with Sensor Settings
The "Fill with Sensor" button of the "Edit User Correction Data" menu opens the asso-
ciated menu.
This dialog describes all parameters for filling a table automatically with sensor read-
ings.
To select the sensor and determine its parameters, refer to
To fill the table, press the "Execute" button.
Fill User Correction Data with Sensor
Enables you to fill the table with correction data acquired by a connected power sensor
from Rohde
&
Schwarz.
"Sensor"
Displays connected sensors for selection.
"List To Fill"
Indicates the used list file.
"Include Zeroing"
Performs a zeroing procedure before acquiring the user correction
data to improve precision. Since during zeroing no signal may be
applied to the sensor, RF is temporarily switched off at the generator.
When unchecked, the zeroing procedure is skipped. The RF signal
level might be blanked shortly. This setting is recommended if blank-
ing of RF is undesirable or the absence of power at the sensor can
not be guaranteed.
"Execute"
Performs automatic filling of the list, provided a sensor is detected
and the user correction list contains at least one frequency value.
RF Level