Instrument Function
R&S
®
SMF100A
225
Operating Manual 1167.2319.02 ─ 12
The start and stop time of the gates are indicated as gate markers, a bar between the
start and stop marker shows the gate length. The indication state of the gate borders
and measurement values is only available for certain diagram views which are
switched with the "REARR" key. The "REARR list NRP-Z Analysis" dialog provides a
selection of views, between which is toggled (see
" REARR list - Power Analysis "
In the remote control mode, commands for setting the indication state differ from the
commands for reading the measurement values.
●
Command that defines the indication state for the diagram and for hardcopy:
:TRACe[:POWer]:SWEep:MEASurement:GATE:DISPlay:ANNotation[:
STATe]
●
Commands that query the measured values:
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:AVERage?
Queries the measured average power.
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:MAXimum?
Queries the measured peak power.
Trace - Gate
Selects the trace to which the gates are assigned. The sensor assignment to the
respective trace is performed in the measurement diagram (trace buttons). The two
gates are assigned to the same trace.
Remote command:
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:FEED
Indication
(time measurement mode only)
RF Level