Instrument Function
R&S
®
SMF100A
207
Operating Manual 1167.2319.02 ─ 12
Refer to the manual of the connected R&S NRP-Zxx power sensor for a description on
how to use the SParameter table.
Remote command:
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
5.4.6.3
NRP-Z Power Analysis
The signal generator in combination with a connected R&S NRP probe allows sweep
measurements on DUTs.
"NRP-Z Power Analysis" automatically disables
Access:
► Select "Level Control > NRP-Z Power Analysis".
The measurement data of the sensors is displayed in traces in a measurement dia-
gram. Four traces are available. The traces are automatically or manually assigned to
the sensors. In addition to the data traces, a reference trace can be stored and recalled
and/or the trace indication can be frozen temporarily (hold trace), thus enabling com-
parison of traces. Readout and comparison of particular values of the traces is possible
by means of four markers.
Three measurement modes are provided:
●
Power versus frequency (frequency response)
●
Power versus power (power sweep, AM/AM)
These two modes are generator driven, that means, the generator provides the
measurement signal.
●
Power versus time (power measurement in the time domain R&S NRP trace
mode).
This mode is signal driven, that means besides the generator signal also external
signals can be analyzed. Time mode requires an additional trigger event, for which
RF Level