Instrument Function
R&S
®
SMF100A
197
Operating Manual 1167.2319.02 ─ 12
NRP-Z Power Viewer Settings
► To access the dialog for configuring the RF signal level, perform one of the follow-
ing:
● Select "Level Control > config... > NRP-Z Power Viewer".
● Press the MENU key and select "Level Control > NRP-Z Power Viewer".
"Sensor 1" is firmly assigned to the sensor on the circled SENSOR socket. If no
sensor is connected to this socket, channel 1 remains not assigned. Sensors 2...4
are assigned to the sensors at the USB connectors, according to their sequence of
connection.
When you connect your power sensor(s) via the R&S NRP-Z5 USB sensor hub, each
channel of the hub is firmly assigned to the associated sensor channel in the genera-
tor.
The remote commands required to define the settings are described in
"Power Sensor Measurement Subsystems"
Sensor
Indicates the connector used for the detected sensors.
In remote control, the sensors are set up using the
SENSe
commands. The remote
measurement is triggered by the
READ
query which also provides the measurement
results.
The sensor is selected by suffix 1, 2, 3 or 4 in key word
SENSe
or
READ
of the com-
mand header.
Suffix 1 denotes the sensor connected to SENSOR, suffix 2 the one at the first USB
interface, and suffix 3 and 4 are assigned to the sensors at the following USB interfa-
ces. The suffix is identical to the index which is assigned automatically to each sensor
upon connection.
In order to detect all connected sensors the state of all four connectors (i.e. SENsor1/
SENSor2/SENSor3/SENSor4) must be checked.
Note:
The software version of the connected power sensor can be retrieved by means
of the remote control command
:SENS:POW:TYPE?
.
RF Level