Instrument Function
R&S
®
SMF100A
189
Operating Manual 1167.2319.02 ─ 12
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single", see
Remote command:
[:SOURce<hw>]:SWEep:POWer:RETRace
Step - Level Sweep
Sets the step width for the individual sweep steps. This entry is effective for all sweep
modes.
With the level sweep, the logarithmic step width is a constant fraction of the current
level. This fraction is added to the current level. The logarithmic step width is entered in
dB.
Remote command:
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic]
Count - RF Level Sweep
Sets the number of sweep steps.
Remote command:
[:SOURce<hw>]:SWEep:POWer:POINts
Dwell Time - Level Sweep
Enters the dwell time and determines the duration of the individual sweep steps.
RF Level