Instrument Function
R&S
®
SMF100A
183
Operating Manual 1167.2319.02 ─ 12
Variation Active
Activates the user-defined step width used when varying the level value with the rotary
knob.
"ON"
The level value set with the rotary knob is varied using the user-
defined step width which is entered under "Variation Step".
"OFF"
The level value set with the rotary knob is varied in steps of one unit
at the cursor position (standard operating mode).
Remote command:
Variation Step
Sets the user-defined step width for entering the RF level using the rotary knob. Level
variation with this step width must also be activated with "Variation Active".
Remote command:
[:SOURce<hw>]:POWer:STEP[:INCRement]
5.4.2.3
External Level Adjustment
The external level adjustment provides information about the data that has been used
for calibrating the RF level.
By default the instrument uses correction data obtained in the factory before delivery.
In exceptional cases, you can determine the calibration values with an R&S NRP
power sensor, and use these values for the external level correction. This feature is a
protected function (see Service Manual, chapter 2, "Adjustment").
Adjustment Data
Indicates what data has been used for level calibration.
Remote command:
:CALibration<hw>:LEVel:EXTern:DATA
5.4.3 Power-On/EMF Settings
The power-on behavior of the R&S
SMF and the level display in the display header are
set in the "Power-On / EMF Settings" section of the "RF Level/EMF/ALC/UCOR" dia-
log.
To open the dialog, select "Level Control > Configure > Level" or use the MENU key
under "Level Control".
Power-On State - RF Signal
Selects the state which the RF output is to assume after the instrument is switched on.
"RF Off"
The output is deactivated when the instrument is switched on.
RF Level