Remote Control Commands
R&S
®
SMC100A
310
Operating Manual 1411.4060.32 ─ 04
SYSTem:ERRor:CODE[:NEXT]?
The command queries the oldest entry in the error queue and then deletes it. Only the
error number is returned and not the entire error text.
Return values:
<Next>
string
0
"No error", i.e. the error queue is empty
positive value
Positive error numbers denote device-specific errors
negative value
Negative error numbers denote error messages defined by SCPI.
Example:
SYST:ERR:CODE
queries the oldest entry in the error queue.
Response: 0
No errors have occurred since the error queue was last read out.
Usage:
Query only
SYSTem:ERRor:COUNt?
The command queries the number of entries in the error queue. If the error queue is
empty, '0' is returned.
Return values:
<Count>
string
Example:
SYST:ERR:COUN
queries the number of entries in the error queue.
Response: 1
One error has occurred since the error queue was last read out.
Usage:
Query only
SYSTem:ERRor[:NEXT]?
The command queries the oldest entry in the error queue and then deletes it.
The command is identical to the command
STATus:QUEue:NEXT
.
Return values:
<Next>
string
0
"No error", i.e. the error queue is empty
positive value
Positive error numbers denote device-specific errors
negative value
Negative error numbers denote error messages defined by SCPI.
SYSTem Subsystem