Remote Control Commands
R&S
®
SMC100A
235
Operating Manual 1411.4060.32 ─ 04
........................................................236
...........................................................236
..........................................................236
....................................................................237
...................................................................237
..............................................................................237
...................................................................238
...............................................................................238
.................................................................................239
...................................................................................239
.........................................................................................239
..............................................................................240
......................................................................................240
.......................................................................................240
...........................................................................241
.....................................................................................241
............................................................................................241
...........................................................................................241
.............................................................................................242
INITiate<ch>[:POWer]:CONTinuous
<Continuous>
The command switches the local state of the continuous power measurement by the R&S
NRP-Zxx power sensors on and off. Switching the local state off enhances the measure-
ment performance during remote control
The remote measurement is triggered by the READ query (command
) which also provides the measurement results. The local state is not influenced
by this command, measurements results can be retrieved with local state on or off.
Parameters:
<Continuous>
0 | 1 | OFF | ON
*RST:
OFF
Example:
INIT:CONT ON
switches local state of continuous power measurement on.
READ<ch>[:POWer]?
The command triggers the measurement with power sensors and provides the power
measurement result of the selected power sensor. The value is provided with the unit set
with command
SENSe:UNIT[:POWer]
.
For certain power sensors, e.g. R&S NRP-Z81, two values are returned, first the value
for the average level and - separated by a comma - the peak level
Note:
The local state is not influenced by this command, measurements results can be
retrieved with local state on or off. For long measurement times it is recommended to use
a SRQ (MAV bit) for command synchronization.
SENSe, READ and INITiate Subsystems
Содержание SMC100A
Страница 1: ...R S SMC100A Signal Generator Operating Manual Operating Manual XÊP 1411 4060 32 04 Test Measurement ...
Страница 67: ......
Страница 339: ......
Страница 341: ......