Remote Control Commands
R&S
®
SMBV100B
876
User Manual 1178.4460.02 ─ 03
Example:
Manual operation:
See
[:SOURce<hw>]:POWer:RANGe:LOWer?
[:SOURce<hw>]:POWer:RANGe:UPPer?
Queries the current interruption-free range of the level.
Return values:
<Upper>
float
Increment: 0.01
Default unit: dBm
Example:
SOURce1:POWer:RANGe:UPPer?
// -15
SOURce1:POWer:RANGe:LOW?
// -50
Usage:
Query only
Manual operation:
See
[:SOURce]:POWer:WIGNore
<State>
Ignores level range warnings.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
SOURce:POWer:WIGNore 1
12.16.16
SOURce:ROSCillator Subsystem
The
SOURce:ROSCillator
subsystem contains the commands for setting the exter-
nal and internal reference frequency.
The commands of this subsystem are not affected by an instrument reset (
SOURce Subsystem