Remote Control Commands
R&S
®
SMBV100B
857
User Manual 1178.4460.02 ─ 03
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also
.
Range:
depends on the installed options
Default unit: dBm
Example:
See
Example "Create an RF list and activate the list mode"
Manual operation:
See
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the number (points) of level entries in the selected list.
Return values:
<Points>
integer
Range:
0 to INT_MAX
*RST:
0
Example:
See
Example "Create an RF list and activate the list mode"
Usage:
Query only
[:SOURce<hw>]:LIST:TRIGger:EXECute
Starts the processing of a list in list mode.
Example:
See
Example "Create an RF list and activate the list mode"
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:LIST:TRIGger:SOURce
<Source>
Selects the trigger source for processing lists.
The names of the parameters correspond to those in sweep mode. SCPI standard
uses other names for the parameters; these names are also accepted by the instru-
ment. The SCPI names should be used if compatibility is an important consideration.
For an overview, see the following table:
SOURce Subsystem