Remote Control Commands
R&S
®
SMBV100B
650
User Manual 1178.4460.02 ─ 03
The local state is set with the
INIT
command. Switching off the local state enhances
the measurement performance. Measurements results can be retrieved in local state
on or off.
Sensor parameters are set with the
SENSe
commands.
To start the measurement and retrieve the result, use the
command.
Suffix
Value range
Description
SENSe<ch>
[1] to 4
Indicates the sensor
Sensor mapping:
●
SENSe[1]
- default mapping for sensors connected to the
[Sensor] connector
●
SENSe2
- sensor connected to a [USB] connector
●
SENSe3|4
- further connected sensors to [USB] connectors,
in the connection order
Use the
:SLISt
commands to change the sensor mapping
READ<ch>
[1] to 4
Sensor assignment
INIate<hw>
[1] to 4
Sensor assignment
ELEMent<ch>
[1] to 25
Sensor mapping list
Programming examples
Example: Detecting and assigning a power sensor
SLISt:LIST?
// Response: "NRP33SN-V-900007-USB Legacy","NRP-Z211-900001-USB Legacy"
// list of automatically detected sensors
SLISt:SCAN:STATe 1
// searches for sensors connected in the LAN or via the USBTMC protocol
SLISt:LIST?
// Response:
// "NRP33SN-V-900007-USB Legacy","NRP-Z211-900001-USB Legacy",
// "NRP33SN-V-900005-USBTMC","NRP33SN-V-900011-LAN"
// list of automatically detected sensors
// the list can contain more entries
SLISt:ELEMent3:MAPPing SENS1
// maps the third sensor from the list to the first sensor channel
SLISt:SENSor:MAP "NRPS18S-100654-USB Legacy", SENS3
// maps the sensor directly to channel 3
SENSe, READ, INITiate and SLISt Subsystems