Parameters:
<Logarithmic>
float
Range:
0.01 to 100
Increment: 1E-3
*RST:
1
Example:
FREQ:STAR 1GHz
sets the start frequency to 1 GHz.
FREQ:STOP 5GHz
sets the stop frequency to 5 GHz.
SWE:SPAC LOG
sets logarithmic sweep spacing.
SWE:STEP:LOG 10PCT
sets the step width for logarithmic sweep spacing to 10% of the
previous frequency in each instance (for a frequency sweep).
Manual operation:
See
" Step Lin/Log - Frequency Sweep "
[:SOURce<hw>]:SWEep:POWer:DATTenuator:STATe
<Dattstate>
Activates digital attenuation during RF power sweep.
Parameters:
<Dattstate>
0 | 1 | OFF | ON
*RST:
0
Example:
SWE:POW:DATT:STAT ON
switches on digital attenuation.
Manual operation:
See
[:SOURce<hw>]:SWEep:POWer:DWELl
<Dwell>
Sets the time taken for each level step of the sweep.
Tip:
It is recommended to switch off the "Display Update" for optimum sweep perform-
ance especially with short dwell times (
Parameters:
<Dwell>
float
Range:
1E-3 to 100
Increment: 100E-6
*RST:
15E-3
Example:
SWE:POW:DWEL 12 ms
sets a dwell time of 12 ms for a level sweep at the RF output.
Manual operation:
See
[:SOURce<hw>]:SWEep:POWer:EXECute
Triggers a sweep.