Remote Control Commands
R&S
®
SMBV100A
767
Operating Manual 1176.8016.02 ─ 17
Example:
LFO:SHAP SQU
selects a rectangular shape for the signal of the LF generator.
Manual operation:
See
[:SOURce]:LFOutput:SIMPedance
<SImpedance>
Selects the output impedance of the LF generator. Selection "LOW" and "600 Ohm"
are available.
Note:
The installed hardware determines the available settings. Use the
dialog to check the hardware the instrument is equipped with.
For information on the required hardware revision, refer to the release notes.
Parameters:
<SImpedance>
LOW | G600
*RST:
LOW
Example:
SOUR:LFO:SIMP G600
'sets the output impedance of the LF generator to 600 Ohms
Manual operation:
See
[:SOURce]:LFOutput:VOLTage
<Voltage>
Sets the voltage of the LF output signal.
Parameters:
<Voltage>
float
Range:
see the data sheet: Internal modulation generator >
Output voltage range
Increment: see the data sheet: resolution of output voltage set-
ting
*RST:
1
Example:
LFO:VOLT 2 V
sets the voltage of the LF output to 2 V.
Manual operation:
See
7.16.12 SOURce:LIST Subsystem
This subsystem contains the commands for the List mode of the instrument.
The following settings are required to operate the instrument in List mode:
1. Create a list.
If a list which does not exist is selected with the
:LIST:SEL
command, an empty
list with the name of the selected list is created.
SOUR1:LIST:SEL "New_list"
SOURce Subsystem